File name calk2000_testver2.txtKeithley Model 2000 - detected, S/N ???????, Version: A20 /A02
Environment temperature = 31.94 C, humidity = 35.08 RH, pressure = 1005.6305 hPa
HP 3458A detected
TEMP? = 37.6 C
Ambient temp = +24.7C
HP 3458A DCV 1.692330780E-06 VDC, K2000 DUT 2.770412910E-06 VDC, Delta -389141.317 ppm
MFC error = 197.687 ppm
HP 3458A DCV 9.999880032E-03 VDC, K2000 DUT 1.000043996E-02 VDC, Delta -55.991 ppm
MFC error = 31.682 ppm
HP 3458A DCV 9.999996367E-02 VDC, K2000 DUT 1.000009003E-01 VDC, Delta -9.367 ppm
MFC error = -12.926 ppm
HP 3458A DCV 9.999993737E-01 VDC, K2000 DUT 9.999957880E-01 VDC, Delta 3.586 ppm
MFC error = -3.706 ppm
HP 3458A DCV 9.999983129E+00 VDC, K2000 DUT 9.999996827E+00 VDC, Delta -1.370 ppm
MFC error = 3.901 ppm
HP 3458A DCV 1.000005225E+02 VDC, K2000 DUT 9.999935250E+01 VDC, Delta 11.700 ppm
MFC error = 19.913 ppm
HP 3458A DCV 1.000003349E+03 VDC, K2000 DUT 1.000007700E+03 VDC, Delta -4.351 ppm
MFC error = 160.557 ppm
Measured 2000 DCI 5 : 3.333360000E+02 ADC
HP 3458A DCV 9.999841548E-05 ADC, K2000 DUT 3.333360000E+02 ADC, Delta -999999.700 ppm
Measured REF DCI 5 : 9.999841548E-05 ADC, stored DCI = 0.099984
MFC error = 6.350 ppm
Measured 2000 DCI 5 : 1.111130000E+02 ADC
Measured REF DCI 5 : 9.999984376E-04 ADC, stored DCI = 0.999994
MFC error = 14.367 ppm
Measured 2000 DCI 5 : 3.704766992E+01 ADC
Measured REF DCI 5 : 9.999987782E-03 ADC, stored DCI = 9.999856
MFC error = -13.773 ppm
Measured 2000 DCI 5 : 1.244922433E+01 ADC
Measured REF DCI 5 : 9.999964881E-02 ADC, stored DCI = 100.001377
MFC error = -77.430 ppm
Measured 2000 DCI 5 : 5.149731650E+00 ADC
Measured REF DCI 5 : 9.999958700E-01 ADC, stored DCI = 1.000077
MFC error = -10000000000000000139372116959414099130712064.000 ppm
MFC error = -1000000000000000044885712678075916785549312.000 ppm
MFC error = -100000000000000000620008645040778319495168.000 ppm
MFC error = -9999999999999999094860208812374492184576.000 ppm |